Three-dimensional numerical simulation of a failed coronary stent implant at different degrees of residual stenosis. Part I: Fluid dynamics and shear stress on the vascular wall

Date

2017-03-28

Advisors

Journal Title

Journal ISSN

Volume Title

Publisher

Taylor and Francis

Department

Type

Article

ISSN

1040-7782

item.page.extent-format

Citation

Di Venuta I, Boghi A, Gori F, Three-dimensional numerical simulation of a failed coronary stent implant at different degrees of residual stenosis. Part I: Fluid dynamics and shear stress on the vascular wall, Numerical heat transfer, Part A: applications, Volume 71, Issue 6, Pages 638 – 652.

Abstract

The influence of the degree of residual stenosis on the hemodynamics inside coronary arteries is investigated through three-dimensional (3D) numerical simulations. The present paper, which is the first of a series of two, focuses on the influence that the degree of residual stenosis (DOR) has on the fluid dynamics and the shear stresses acting on the stent and the artery wall. The pulsatile nature of the blood flow and its non-Newtonian features are taken into account. Four models of artery are investigated. The results show that the wall shear stress (WSS) increases monotonically, but not linearly, with the DOR.

Description

item.page.description-software

item.page.type-software-language

item.page.identifier-giturl

Keywords

stent, stenosis, wall shear stress, New-Newtonian

Rights

Attribution-NonCommercial 4.0 International

item.page.relationships

item.page.relationships

item.page.relation-supplements