Development of baseline stability in an electromagnetic flowmeter for dielectric liquids

dc.contributor.advisorHemp, J.
dc.contributor.advisorSanderson, M. L.
dc.contributor.authorDurcan, L. P.
dc.date.accessioned2016-10-13T08:59:22Z
dc.date.available2016-10-13T08:59:22Z
dc.date.issued1998-10
dc.description.abstractThis thesis is concerned with the development of a prototype electromagnetic flowmeter for dielectric liquids. A initial prototype flowmeter is tested in laboratory conditions with a liquid of low conductivity, the conclusion of these initial tests being that the stability of the flowmeter is below that required for its design applications. This lack of stability is a problem reported on several occasions in similar projects involving electromagnetic flowmeters for dielectric liquids. Therefore the main body of the work is spent investigating the causes of instability and trying to redress the necessary problems in order that a viable working prototype is developed. Problems of thermal instability, electrostatic and electromagnetic shielding are addressed along with the causes and effects of all the major noise and unwanted signals experienced in such a system. A new prototype is developed incorporating stability driven improvements. This new design is rigorously tested and assessed. The stability of the meter is considered with respect to its flow related output. Several parameters of operational stability are given in the examples considered. A final assessment is then made of the flowmeters baseline performance and a account of any remaining instability is given with respect to the issue developed throughout the thesis. The conclusions indicate that the electromagnetic flowmeter for dielectric fluids does produce a linear output based on Faraday's law of induction. They also indicate that the instability experienced previously can be broken into its constituent parts and each contributor accounted for and reduced, working towards the ideal totally stable flowmeter.en_UK
dc.identifier.urihttp://dspace.lib.cranfield.ac.uk/handle/1826/10720
dc.language.isoenen_UK
dc.publisherCranfield Universityen_UK
dc.rights© Cranfield University, 1998. All rights reserved. No part of this publication may be reproduced without the written permission of the copyright holder.en_UK
dc.titleDevelopment of baseline stability in an electromagnetic flowmeter for dielectric liquidsen_UK
dc.typeThesis or dissertationen_UK
dc.type.qualificationlevelDoctoralen_UK
dc.type.qualificationnamePhDen_UK

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
Durcan_L_P_1998.zip
Size:
21.65 MB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.79 KB
Format:
Item-specific license agreed upon to submission
Description: