Amperometric biosensor for formic acid in air
Date published
2000-11-01
Free to read from
Supervisor/s
Journal Title
Journal ISSN
Volume Title
Publisher
Elsevier
Department
Type
Article
ISSN
0925-4005
Format
Citation
K. J. Mattias Sandström, Jeffrey Newman, Anna-Lena Sunesson, Jan-Olof Levina, and Anthony P. F. Turner, Amperometric biosensor for formic acid in air, Sensors and Actuators B, Volume 70, Issues 1-3, 1 November 2000, Pages 182-187
Abstract
The possibility of developing a simple, inexpensive and specific personal passive ”real-time” air sampler based on biosensor technology was investigated. Formic acid was used as a model substance. The sensor is based on the enzymatic reaction between formic acid and formate dehydrogenase with NAD+ as a cofactor and Meldola’s blue as mediator. An effective way to immobilise the enzyme, cofactor and Meldola’s blue on screen-printed electrodes was found to be in a mixture of glycerol and phosphate buffer covered with a gas-permeable membrane. When the sensor was introduced into an atmosphere containing formic acid, it gave a distinct and rapid amperometric response.