Highly sensitive contact pressure measurements using FBG patch in endotracheal tube cuff
dc.contributor.author | Correia, Ricardo N. | |
dc.contributor.author | Blackman, O. R. | |
dc.contributor.author | Hernandez, F. U. | |
dc.contributor.author | Korposh, Sergiy | |
dc.contributor.author | Morgan, S. P. | |
dc.contributor.author | Hayes-Gill, R. | |
dc.contributor.author | James, Stephen W. | |
dc.contributor.author | Evans, D. | |
dc.contributor.author | Norris, A. | |
dc.date.accessioned | 2016-07-05T10:01:14Z | |
dc.date.available | 2016-07-05T10:01:14Z | |
dc.date.issued | 2016-05-31 | |
dc.description.abstract | A method for measuring the contact pressure between an endotracheal tube cuff and the trachea was designed and developed by using a fibre Bragg grating (FBG) based optical fibre sensor. The FBG sensor is encased in an epoxy based UV-cured cuboid patch and transduces the transversely loaded pressure into an axial strain that induces wavelength shift of the Bragg reflection. The polymer patch was created by using a PTFE based mould and increases tensile strength and sensitivity of the bare fibre FBG to pressure to 2.10×10-2 nm/kPa. The characteristics of the FBG patch allow for continuous measurement of contact pressure. The measurement of contact pressure was demonstrated by the use of a 3D printed model of a human trachea. The influence of temperature on the measurements is reduced significantly by the use of a second FBG sensor patch that is not in contact with the trachea. Intracuff pressure measurements performed using a commercial manometer agreed well with the FBG contact pressure measurements. © (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only. | en_UK |
dc.identifier.citation | Correia et al. (2016) Highly sensitive contact pressure measurements using FBG patch in endotracheal tube cuff, 6th European Workshop on Optical Fibre Sensors (EWOFS'2016), Limerick, Ireland, 30-31 May 2016 | en_UK |
dc.identifier.uri | http://dx.doi.org/10.1117/12.2236859 | |
dc.identifier.uri | https://dspace.lib.cranfield.ac.uk/handle/1826/10081 | |
dc.language.iso | en | en_UK |
dc.publisher | SPIE | en_UK |
dc.rights | Attribution-NonCommercial-NoDerivatives 4.0 International | en_UK |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/ | |
dc.subject | Fibre Bragg Grating | en_UK |
dc.subject | FBG | en_UK |
dc.subject | Intracuff pressure | en_UK |
dc.subject | Contact pressure | en_UK |
dc.subject | Endotracheal tube | en_UK |
dc.subject | ETT | en_UK |
dc.title | Highly sensitive contact pressure measurements using FBG patch in endotracheal tube cuff | en_UK |
dc.type | Article | en_UK |
Files
Original bundle
1 - 1 of 1
Loading...
- Name:
- Highly_sensitive_contact_pressure_measurements_using_FBG_patch_in_endotracheal_tube_cuff_2016.pdf
- Size:
- 518.33 KB
- Format:
- Adobe Portable Document Format
License bundle
1 - 1 of 1
No Thumbnail Available
- Name:
- license.txt
- Size:
- 1.79 KB
- Format:
- Item-specific license agreed upon to submission
- Description: