Highly sensitive contact pressure measurements using FBG patch in endotracheal tube cuff

Citation

Correia et al. (2016) Highly sensitive contact pressure measurements using FBG patch in endotracheal tube cuff, 6th European Workshop on Optical Fibre Sensors (EWOFS'2016), Limerick, Ireland, 30-31 May 2016

Abstract

A method for measuring the contact pressure between an endotracheal tube cuff and the trachea was designed and developed by using a fibre Bragg grating (FBG) based optical fibre sensor. The FBG sensor is encased in an epoxy based UV-cured cuboid patch and transduces the transversely loaded pressure into an axial strain that induces wavelength shift of the Bragg reflection. The polymer patch was created by using a PTFE based mould and increases tensile strength and sensitivity of the bare fibre FBG to pressure to 2.10×10-2 nm/kPa. The characteristics of the FBG patch allow for continuous measurement of contact pressure. The measurement of contact pressure was demonstrated by the use of a 3D printed model of a human trachea. The influence of temperature on the measurements is reduced significantly by the use of a second FBG sensor patch that is not in contact with the trachea. Intracuff pressure measurements performed using a commercial manometer agreed well with the FBG contact pressure measurements. © (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.

Description

item.page.description-software

item.page.type-software-language

item.page.identifier-giturl

Keywords

Fibre Bragg Grating, FBG, Intracuff pressure, Contact pressure, Endotracheal tube, ETT

Rights

Attribution-NonCommercial-NoDerivatives 4.0 International

item.page.relationships

item.page.relationships

item.page.relation-supplements