Finite element analysis of SS316L-based five-hole plate implant for fibula reconstruction

Date

2022-08-15

Supervisor/s

Journal Title

Journal ISSN

Volume Title

Publisher

Intellectual Research and Publication Foundation (Yayasan Riset dan Publikasi Intelektual)

Department

Type

Article

ISSN

2715-6087

Format

Citation

Qosim N, Emzain ZF, Mufarrih AM, et al., (2022) Finite element analysis of SS316L-based five-hole plate implant for fibula reconstruction. Journal of Applied Engineering and Technological Science (JAETS), Volume 4, Issue 1, August 2022, pp. 16-23

Abstract

This study analyzed the design performance of SS316L-based plate implant for fibula restoration using a Finite Element Analysis approach. The simulated model design has dimensions of 35 x 5 x 1.5 mm and five holes with 2-3 configuration. The results of the bending test simulation showed that the values for both displacement and Von Mises stress that occurred (0.008 mm and 116 MPa of each) were still considerably below the yield stress of the SS316L material. The same results were also shown in the tensile test simulation, although the clamping setting on the plate was changed on the other side. From this finite element analysis approach, the SS316L-based five-hole plate implant design has a fairly good strength performance as a fibular bone-implant restoration product.

Description

Software Description

Software Language

Github

Keywords

plate, implant, finite element method, von mises, fibula, SS316L

DOI

Rights

Attribution 4.0 International

Relationships

Relationships

Supplements